Back to [[FDG-PET]]
 
Back to [[FDG-PET]]
  
==QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee Charge==
+
==QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee==
  
The '''objective of this subcommittee''' is to build a common reference DICOM test image that can be generated by each scanner and read on PET DICOM display stations to check information fidelity (i.e. next step beyond DICOM conformance statement)
+
The '''objective of this subcommittee''' is to define how to calculate SUV in a consistent manner regardless of manufacturer, model and version of scanner and workstation.
  
 
'''Completed:'''
 
'''Completed:'''
 
*Collected DICOM PET/CT images of the same reference phantom from scanners from GE, Philips, Siemens
 
*Collected DICOM PET/CT images of the same reference phantom from scanners from GE, Philips, Siemens
 
*Image data collated and compared to DICOM header information
 
*Image data collated and compared to DICOM header information
 +
*Collected answers from vendors to questionnaires along with other submitted SUV calculation documents
 +
*Described generalized SUV calculation method with vendor-specific variants
  
==Working Documents==
+
The previously described objective of building a common reference DICOM test image that can be generated by each scanner and read on PET DICOM display stations to check information fidelity is the subject of the separate [[University of Washington]] Digital Reference Object (DRO) project (described in the [[Media:12-month_12a_report_QIBA_FDG-PETCT_DRO_final_report_v3.pdf|12-Month Final Project Report]]).
*[[Media:Dr-Kinahan-DRO-20090310.ppt|Digital Reference Objects - 4 Slide Overview - 2009.03.10]]
+
 
*[[Media:SUV_vendorneutral_pseudocode_20120928_DAC.doc|Vendor-neutral pseudo-code for SUV calculation - updated 20120928 - changes from 20091106 tracked]]
+
==SUV Calculation==
 +
*[[Media:SUV_vendorneutral_pseudocode_20180626_DAC.pdf|Vendor-neutral pseudo-code for SUV calculation]] - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time
 +
*[[Media:SUV_vendorneutral_pseudocode_happypathonly_20180626_DAC.pdf|Vendor-neutral pseudo-code for SUV calculation - extracted "happy path only"]] - for clarity (BQML, no private data elements required, series time is OK) - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time
  
 
== Letter to vendors requesting detailed SUV computation description ==
 
== Letter to vendors requesting detailed SUV computation description ==
Line 23: Line 27:
 
* Philips - [[Media:QIBA_FDG_Vendor_SUV_Philips.doc]]
 
* Philips - [[Media:QIBA_FDG_Vendor_SUV_Philips.doc]]
  
* Siemens - [[Media:QIBA_FDG_Vendor_SUV_20081103_DAC_Siemens.doc]]
+
* Siemens - [[Media:QIBA_FDG_Vendor_SUV_20081103_DAC_Siemens.doc]] [[Media:BiographDecayCorr.doc]]
  
 
* GE - [[Media:Calculation_of_SUVs_in_GE_Apps_v4_(2).doc]]
 
* GE - [[Media:Calculation_of_SUVs_in_GE_Apps_v4_(2).doc]]
  
 
* MIMvista - [[Media:MIMvista_QIBA_FDG_Vendor_SUV_20081103_DAC-2.doc‎]]
 
* MIMvista - [[Media:MIMvista_QIBA_FDG_Vendor_SUV_20081103_DAC-2.doc‎]]
 
== Vendor supplied information ==
 
* GE
 
* Siemens - [[Media:BiographDecayCorr.doc]]
 
  
 
==Subcommittee Call Summaries==
 
==Subcommittee Call Summaries==
 
+
'''SUV'''
 
*[[Media:11.05.09 QIBA Q-PET-SUV SubCtte Call Summary-FINAL.pdf|11.05.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 
*[[Media:11.05.09 QIBA Q-PET-SUV SubCtte Call Summary-FINAL.pdf|11.05.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 
*[[Media:09.30.2009 QIBA FDG-PET CT SUV SubCtts Call Summary-FINAL.pdf|09.30.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 
*[[Media:09.30.2009 QIBA FDG-PET CT SUV SubCtts Call Summary-FINAL.pdf|09.30.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 
*[[Media:08.06.2009 QIBA FDG-PET SUV Subctte Update Call-FINAL.pdf|08.06.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 
*[[Media:08.06.2009 QIBA FDG-PET SUV Subctte Update Call-FINAL.pdf|08.06.2009 FDG-PET/CT SUV Subcommittee Call Summary]]
 +
'''DRO'''
 
*[[Media:04.17.2009 QIBA FDG-PET DRO Subctte Call Summary-FINAL.pdf|04.17.2009 FDG-PET/CT DRO Subcommittee Call Summary]]
 
*[[Media:04.17.2009 QIBA FDG-PET DRO Subctte Call Summary-FINAL.pdf|04.17.2009 FDG-PET/CT DRO Subcommittee Call Summary]]
 
*[[Media:03.27.2009 QIBA FDG-PET DRO Call Summary-FINAL.pdf|03.27.2009 FDG-PET/CT DRO Subcommittee Call Summary]]
 
*[[Media:03.27.2009 QIBA FDG-PET DRO Call Summary-FINAL.pdf|03.27.2009 FDG-PET/CT DRO Subcommittee Call Summary]]

Latest revision as of 11:48, 31 December 2021

Back to FDG-PET

QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee

The objective of this subcommittee is to define how to calculate SUV in a consistent manner regardless of manufacturer, model and version of scanner and workstation.

Completed:

  • Collected DICOM PET/CT images of the same reference phantom from scanners from GE, Philips, Siemens
  • Image data collated and compared to DICOM header information
  • Collected answers from vendors to questionnaires along with other submitted SUV calculation documents
  • Described generalized SUV calculation method with vendor-specific variants

The previously described objective of building a common reference DICOM test image that can be generated by each scanner and read on PET DICOM display stations to check information fidelity is the subject of the separate University of Washington Digital Reference Object (DRO) project (described in the 12-Month Final Project Report).

SUV Calculation

Letter to vendors requesting detailed SUV computation description

Vendor responses on SUV computation

Subcommittee Call Summaries

SUV

DRO

SUV Subcommittee Members

  • Paul Kinahan, PhD, Subcommittee Co-chair <kinahan@u.washington.edu>
  • David Clunie, MBBS, Subcommittee Co-chair <dclunie@radpharm.com>
  • Ronald Boellaard, PhD <r.boellaard@vumc.nl>
  • Michael Casey, PhD <michael.e.casey@siemens.com>
  • Paul E. Christian <Paul.Christian@hci.utah.edu>
  • Patricia E. Cole, PhD, MD <patriciaellencole@gmail.com>
  • Simon DeBruin, MSEE <sdebruin@segamicorp.com>
  • John M. Hoffman, MD <john.hoffman@hci.utah.edu>
  • Steve Kohlmyer <Steve.Kohlmyer@med.ge.com>
  • Piotr Maniawski <piotr.maniawski@philips.com>
  • Dennis Nelson, PhD <dnelson@mimvista.com>
  • Eric S. Perlman, MD <perlman@radpharm.com>
  • Ling X. Shao, PhD, <ling.shao@philips.com>
  • John G. Wolodzko, PhD <wolodzko@radpharm.com>
  • Jeffrey Yap, PhD <Jeffrey_Yap@dfci.harvard.edu>
  • Brian Zimmerman, PhD <bez@nist.gov>