+ | == Generic dynamic protocol == | ||
+ | 1.5 Tesla | ||
+ | 3D spoiled gradient echo | ||
+ | Body XMT | ||
+ | Phased array RCV | ||
+ | Quiet breathing | ||
+ | No parallel imaging | ||
+ | No magnetization preparation | ||
+ | Oblique coronal | ||
+ | Frequency encode S/I | ||
+ | TE as short as possible (1-3 ms) | ||
+ | TR as short as possible (5-7 ms) | ||
+ | Temporal resolution ≤10 sec | ||
+ | SNR/DR trade-off flip angle (30 degree) | ||
+ | ±31.25 kHz receiver bandwidth (i.e., 250 Hz/pixel) | ||
+ | Region appropriate FOV (pilot 40 cm for body) | ||
+ | 80% FOV for acquisition (may need to adjust arm position to avoid wrap-around) | ||
+ | Partial Fourier as needed | ||
+ | As many slices per acquisition as possible (~12 prior to zero fill) | ||
+ | 5-8 mm slice thickness | ||
+ | 256 X 160 acquisition matrix (i.e., 1.56 mm × 2.00 mm pixel size) | ||
+ | Zero fill in slice direction (ZIP2) | ||
+ | [For phantom, also zero fill in plane (ZIP512)] | ||
+ | Acquire data for 10 minutes, inject at 1 minute | ||
+ | |||
+ | == Generic ratio protocol == | ||
+ | These images will be acquired prior to contrast injection | ||
+ | All parameters the same as for dynamic protocol except: | ||
+ | 15 degree flip angle | ||
+ | 8 NEX (averages) | ||
+ | Repeat second time with body RCV | ||
+ | |||
+ | |||
+ | == Generic T1 protocol == | ||
+ | These images will be acquired prior to contrast injection | ||
+ | All parameters the same as for dynamic protocol except: | ||
+ | 3, 6, 9, 15, 24, 35 degree flip angles (may need to be adjusted | ||
+ | – look-up table dependent on TR and scanner) | ||
+ | 4 NEX | ||
+ | |||
+ | |||
Back to [[DCE-MRI]] | Back to [[DCE-MRI]] |
Revision as of 17:12, 6 October 2008
Generic dynamic protocol
1.5 Tesla 3D spoiled gradient echo Body XMT Phased array RCV Quiet breathing No parallel imaging No magnetization preparation Oblique coronal Frequency encode S/I TE as short as possible (1-3 ms) TR as short as possible (5-7 ms) Temporal resolution ≤10 sec SNR/DR trade-off flip angle (30 degree) ±31.25 kHz receiver bandwidth (i.e., 250 Hz/pixel) Region appropriate FOV (pilot 40 cm for body) 80% FOV for acquisition (may need to adjust arm position to avoid wrap-around) Partial Fourier as needed As many slices per acquisition as possible (~12 prior to zero fill) 5-8 mm slice thickness 256 X 160 acquisition matrix (i.e., 1.56 mm × 2.00 mm pixel size) Zero fill in slice direction (ZIP2) [For phantom, also zero fill in plane (ZIP512)] Acquire data for 10 minutes, inject at 1 minute
Generic ratio protocol
These images will be acquired prior to contrast injection All parameters the same as for dynamic protocol except: 15 degree flip angle 8 NEX (averages) Repeat second time with body RCV
Generic T1 protocol
These images will be acquired prior to contrast injection All parameters the same as for dynamic protocol except: 3, 6, 9, 15, 24, 35 degree flip angles (may need to be adjusted – look-up table dependent on TR and scanner) 4 NEX
Back to DCE-MRI