MR Acquisition Protocols: Difference between revisions
Jump to navigation
Jump to search
New page: Back to DCE-MRI |
No edit summary |
||
| Line 1: | Line 1: | ||
== Generic dynamic protocol == | |||
1.5 Tesla | |||
3D spoiled gradient echo | |||
Body XMT | |||
Phased array RCV | |||
Quiet breathing | |||
No parallel imaging | |||
No magnetization preparation | |||
Oblique coronal | |||
Frequency encode S/I | |||
TE as short as possible (1-3 ms) | |||
TR as short as possible (5-7 ms) | |||
Temporal resolution ≤10 sec | |||
SNR/DR trade-off flip angle (30 degree) | |||
±31.25 kHz receiver bandwidth (i.e., 250 Hz/pixel) | |||
Region appropriate FOV (pilot 40 cm for body) | |||
80% FOV for acquisition (may need to adjust arm position to avoid wrap-around) | |||
Partial Fourier as needed | |||
As many slices per acquisition as possible (~12 prior to zero fill) | |||
5-8 mm slice thickness | |||
256 X 160 acquisition matrix (i.e., 1.56 mm × 2.00 mm pixel size) | |||
Zero fill in slice direction (ZIP2) | |||
[For phantom, also zero fill in plane (ZIP512)] | |||
Acquire data for 10 minutes, inject at 1 minute | |||
== Generic ratio protocol == | |||
These images will be acquired prior to contrast injection | |||
All parameters the same as for dynamic protocol except: | |||
15 degree flip angle | |||
8 NEX (averages) | |||
Repeat second time with body RCV | |||
== Generic T1 protocol == | |||
These images will be acquired prior to contrast injection | |||
All parameters the same as for dynamic protocol except: | |||
3, 6, 9, 15, 24, 35 degree flip angles (may need to be adjusted | |||
– look-up table dependent on TR and scanner) | |||
4 NEX | |||
Back to [[DCE-MRI]] | Back to [[DCE-MRI]] | ||
Revision as of 17:12, 6 October 2008
Generic dynamic protocol
1.5 Tesla 3D spoiled gradient echo Body XMT Phased array RCV Quiet breathing No parallel imaging No magnetization preparation Oblique coronal Frequency encode S/I TE as short as possible (1-3 ms) TR as short as possible (5-7 ms) Temporal resolution ≤10 sec SNR/DR trade-off flip angle (30 degree) ±31.25 kHz receiver bandwidth (i.e., 250 Hz/pixel) Region appropriate FOV (pilot 40 cm for body) 80% FOV for acquisition (may need to adjust arm position to avoid wrap-around) Partial Fourier as needed As many slices per acquisition as possible (~12 prior to zero fill) 5-8 mm slice thickness 256 X 160 acquisition matrix (i.e., 1.56 mm × 2.00 mm pixel size) Zero fill in slice direction (ZIP2) [For phantom, also zero fill in plane (ZIP512)] Acquire data for 10 minutes, inject at 1 minute
Generic ratio protocol
These images will be acquired prior to contrast injection All parameters the same as for dynamic protocol except: 15 degree flip angle 8 NEX (averages) Repeat second time with body RCV
Generic T1 protocol
These images will be acquired prior to contrast injection
All parameters the same as for dynamic protocol except:
3, 6, 9, 15, 24, 35 degree flip angles (may need to be adjusted
– look-up table dependent on TR and scanner)
4 NEX
Back to DCE-MRI