==QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee== | ==QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee== | ||
− | The '''objective of this subcommittee''' is to define how to calculate SUV in a consistent manner | + | The '''objective of this subcommittee''' is to define how to calculate SUV in a consistent manner regardless of manufacturer, model and version of scanner and workstation. |
'''Completed:''' | '''Completed:''' | ||
Line 14: | Line 14: | ||
==SUV Calculation== | ==SUV Calculation== | ||
− | *[[Media: | + | *[[Media:SUV_vendorneutral_pseudocode_20180626_DAC.pdf|Vendor-neutral pseudo-code for SUV calculation]] - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time |
− | *[[Media: | + | *[[Media:SUV_vendorneutral_pseudocode_happypathonly_20180626_DAC.pdf|Vendor-neutral pseudo-code for SUV calculation - extracted "happy path only"]] - for clarity (BQML, no private data elements required, series time is OK) - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time |
== Letter to vendors requesting detailed SUV computation description == | == Letter to vendors requesting detailed SUV computation description == |
Latest revision as of 11:48, 31 December 2021
Back to FDG-PET
QIBA FDG-PET/CT Standardized Uptake Value (SUV) Technical Subcommittee
The objective of this subcommittee is to define how to calculate SUV in a consistent manner regardless of manufacturer, model and version of scanner and workstation.
Completed:
- Collected DICOM PET/CT images of the same reference phantom from scanners from GE, Philips, Siemens
- Image data collated and compared to DICOM header information
- Collected answers from vendors to questionnaires along with other submitted SUV calculation documents
- Described generalized SUV calculation method with vendor-specific variants
The previously described objective of building a common reference DICOM test image that can be generated by each scanner and read on PET DICOM display stations to check information fidelity is the subject of the separate University of Washington Digital Reference Object (DRO) project (described in the 12-Month Final Project Report).
SUV Calculation
- Vendor-neutral pseudo-code for SUV calculation - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time
- Vendor-neutral pseudo-code for SUV calculation - extracted "happy path only" - for clarity (BQML, no private data elements required, series time is OK) - updated 20170920 to correct swapped Rescale Slope/Intercept; updated 20180626 to correct DECY term and incorporate start datetime rather than just time
Letter to vendors requesting detailed SUV computation description
- Media:QIBA_FDG_Vendor_SUV_20081103_DAC.doc Letter sent to vendors
- Media:QIBA_FDG_Vendor_SUV_20081103_DAC_JTY.doc Letter with additional questions from Jeff Yap
Vendor responses on SUV computation
- Philips - Media:QIBA_FDG_Vendor_SUV_Philips.doc
Subcommittee Call Summaries
SUV
- 11.05.2009 FDG-PET/CT SUV Subcommittee Call Summary
- 09.30.2009 FDG-PET/CT SUV Subcommittee Call Summary
- 08.06.2009 FDG-PET/CT SUV Subcommittee Call Summary
DRO
- 04.17.2009 FDG-PET/CT DRO Subcommittee Call Summary
- 03.27.2009 FDG-PET/CT DRO Subcommittee Call Summary
- 03.13.2009 FDG-PET/CT DRO Subcommittee Call Summary
SUV Subcommittee Members
- Paul Kinahan, PhD, Subcommittee Co-chair <kinahan@u.washington.edu>
- David Clunie, MBBS, Subcommittee Co-chair <dclunie@radpharm.com>
- Ronald Boellaard, PhD <r.boellaard@vumc.nl>
- Michael Casey, PhD <michael.e.casey@siemens.com>
- Paul E. Christian <Paul.Christian@hci.utah.edu>
- Patricia E. Cole, PhD, MD <patriciaellencole@gmail.com>
- Simon DeBruin, MSEE <sdebruin@segamicorp.com>
- John M. Hoffman, MD <john.hoffman@hci.utah.edu>
- Steve Kohlmyer <Steve.Kohlmyer@med.ge.com>
- Piotr Maniawski <piotr.maniawski@philips.com>
- Dennis Nelson, PhD <dnelson@mimvista.com>
- Eric S. Perlman, MD <perlman@radpharm.com>
- Ling X. Shao, PhD, <ling.shao@philips.com>
- John G. Wolodzko, PhD <wolodzko@radpharm.com>
- Jeffrey Yap, PhD <Jeffrey_Yap@dfci.harvard.edu>