MR Acquisition Protocols

From QIBA Wiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Generic dynamic protocol

1.5 Tesla
3D spoiled gradient echo
Body XMT
Phased array RCV
Quiet breathing
No parallel imaging
No magnetization preparation
Oblique coronal
Frequency encode S/I
TE as short as possible (1-3 ms)
TR as short as possible (5-7 ms)
Temporal resolution ≤10 sec
SNR/DR trade-off flip angle (30 degree)
±31.25 kHz receiver bandwidth (i.e., 250 Hz/pixel)
Region appropriate FOV (pilot 40 cm for body)
80% FOV for acquisition (may need to adjust arm position to avoid wrap-around)
Partial Fourier as needed
As many slices per acquisition as possible (~12 prior to zero fill)
5-8 mm slice thickness
256 X 160 acquisition matrix (i.e., 1.56 mm × 2.00 mm pixel size)
Zero fill in slice direction (ZIP2)
[For phantom, also zero fill in plane (ZIP512)]
Acquire data for 10 minutes, inject at 1 minute

Generic ratio protocol

These images will be acquired prior to contrast injection
All parameters the same as for dynamic protocol except:
  15 degree flip angle
  8 NEX (averages)
  Repeat second time with body RCV


Generic T1 protocol

These images will be acquired prior to contrast injection
All parameters the same as for dynamic protocol except:
  3, 6, 9, 15, 24, 35 degree flip angles (may need to be adjusted
    – look-up table dependent on TR and scanner)
  4 NEX


Back to DCE-MRI